Patient and Operating Room Safety
1-4 RapidVac™ Smoke Evacuator User’s Guide
Power
This unit is equipped from the factory with either a 110 VAC hospital grade
NEMA 5-15 power cord or a 220 VAC CEE7/7 power cord. Should the AC
power cord need replacing to match another plug configuration, the replacement
plug/cable/receptacle configuration must meet or exceed the following
specifications.
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft. (5m)
Plug - minimum 10A 125 VAC
Unit receptacle - IEC female, minimum 10A 125 VAC
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15 ft. (5m)
Plug - minimum 6A 250 VAC
Unit receptacle - IEC female, minimum 6A 250 VAC
Operating Parameters
Transport and Storage
Patient and Operating Room Safety
General
Ambient temperature range 10° C to 40° C (50° F to 104° F)
Relative humidity 30% to 75%, noncondensing
Ambient temperature range -34° C to 65° C (-29° F to 149° F)
Relative humidity 25% to 85%, noncondensing
Warning
Only trained personnel are to operate the RapidVac Smoke Evacuator. Operators
should avail themselves of preclinical training, a review of pertinent literature, and
other appropriate education before attempting to use the RapidVac.
During initial set up, inspect the smoke evacuator for any damage that may have
been caused during shipping. If damaged, do not use or attempt to repair. Call
Valleylab for service assistance.