Cleaning the Smoke Evacuator
After Surgery
RapidVac™ Smoke Evacuator User’s Guide 5-3
3. Disconnect tubing as directed in the previous section.
4. Thoroughly wipe all external surfaces with a mild detergent or disinfectant.
5. Thoroughly clean both sides of the flap covering the filter ports. Do not spray
liquid cleaner into the filter ports.
6. Clean the surface around the filter ports. Do not spray liquid cleaner into the
filter ports. Follow procedures approved by your institution, or use a validated
infection-control procedure.
7. Ensure the smoke evacuator is completely dry before use.
8. Reconnect the power cord to the wall receptacle.
Warning
Wear protective gloves when cleaning around filter ports. Do not place fingers
into ports or touch the internal surface of the ULPA filter. Contaminants may be
present.