23
P/N VT-108723-US, Manual revision: C Vante
®
SEBRA
®
MINI
Tube Sealer Power Source
Table 5-1, Troubleshooting guide
Amber indicator light
on battery charger
fails to illuminate.
Battery pack improperly
connected to charger.
Be sure that the battery
pack is properly
connected to the
charger.
Battery is dead. Replace battery.
Battery charger
improperly connected to
AC power source.
Be sure the battery
charger is connected to
the proper AC power
source.
Note: Arcing at the sealing head may cause a battery low condition to be
indicated, even when the battery is not low.
Failure isolation
If the Sealer fails to operate properly, connect a different sealing head and RF
power cable to the power source to isolate the failed component.
If the Sealer performs properly, then the questionable sealing head is probably
defective. If the Sealer fails to function properly, then the power source may be
defective. In any event, contact the local Vante representative.
CUSTOMER SERVICE
Services
Clinical training
The local Vante representative will provide staff training upon delivery
of the
equipment and should be contacted to organize further instruction if
needed.
Repair service
Vante maintains a worldwide network of company-trained service
representatives responsible for responding to technical needs concerning
equipment. If service beyond the routine maintenance and cleaning described
in this manual is required, the local Vante representative should be
contacted
to provide specific instruction.
Product return guidelines
If, for any reason, merchandise must be returned to the company, the customer
should contact the local Vante representative to arrange for repairs or
returns
using procedures to ensure proper handling and subsequent analysis.
No
returns will be accepted without advanced authorization.
Warning: Vante products must be properly cleaned and packaged
prior
to their return. It remains an important responsibility of the
customer to
reduce potential health hazards by being aware of the risks
involved in
the shipping, handling and testing of this material. Units
returned to
Vante for repair are subject to biohazard charges if
any component is
contaminated with blood or blood products.