EasyManua.ls Logo

Veeco MS40 - Page 8

Veeco MS40
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VIC Leak Detectio
n
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
To fill the pump with oil, unscrew the black cap to the right of the exhaust
filter. Using the oil (1990-924-00) and the funnel (1331-191-00) provided
in the installation kit, slowly fill the pump until the level in the sight glass
is approximately 1/4” below the fill line. Replace the black oil fill cap.
When all shipping material has been removed, all pump connections have
been made and the pump has been filled with oil, replace the right side
panel and close the front and rear doors.
NOTE:
The MS-40 Leak Detector is factory tested and calibrated
prior to shipment and once installed, is ready for use.
An Installation Kit (VIC P/N 0137-800-00) is included with the unit when it
is shipped. The kit contains:
Part Number Description
1680-055-00 Exhaust Filter
0103-141-01 Filament Kit (box of 5)
1331-191-00 Funnel
1330-391-00 7/64 Allen Wrench
1990-924-00 VMP-19 Oil (1 liter)
0137-807-00 Kit, Cart Mounting
0137-804-00 N2 Kit
1621-224-02 Adapter, KF16 to 1/4” NPT
1621-227-05 QC Adapter, KF25 to 1 1/8”
0126-284-00 Blank Plug, 1/18
1620-731-00 Elbow, 1/4 NPT to 3/8 Tube
1750-054-00 Tubing, Tygon, 1/4”
1109-186-00 Rev. C 1-5