User Guide
VentMed
with the following
symbol:
Note 1: At 80MHz, the higher frequency range applied.
Note 2: These guidelines may not be applied in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the device is used exceeds the applicable RF
compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.
b. Over the frequency range 150kHz to 80MHz, the field strengths should be less
than 3 V/m.
Maximum flow rate
1. CPAP
Pressure(cmH2O) Maximum flow rate(L/Min)
4
8
77
111
140
12
38