CPAP, Auto mode for DS-6 Auto CPAP.
CPAP, Auto, S, T, ST mode for DS-7 BIPAP ST25.
CPAP, Auto, S, T, ST, APCV mode for DS-8 BIPAP ST30.
To Unlock Woke Mode parameters
1. Work Mode parameters will be locked automatically after setting in 15
minutes.
2. Long press enter button over 5 seconds to highlight the column to unlock.
To adjust the working pressure
To adjust the working pressure
1. In work mode CPAP, press enter button to highlight Pressure. Press up/down
button to adjust working pressure. You can set pressure 4-20 cmH2O for CPAP