Vernacare Compact Technical Manual-Revision 3-Issued by N.Shields 11.2018
Page | 15
4.2. Siting and fixing
The Vernacare Compact disposal unit should be positioned adjacent to the following:
a) A suitable power supply point.
b) An adequate water supply line.
c) A suitable drainage connection.
The macine must be securely fixed using the recommended installation methods as
shown in section 4.1.
Ensure that adequate space is maintained at each side of the Macerator
to allow for access. Vernacare recommend a gap of 150 - 200mm per side and a
minimum of 50mm away from the wall at the back.
Check that the unit is level in both directions.
230 Volt Machine 10 Amp MCB to BS EN60898 Type C or alternatively fused 10 Amp
to I.E.C 60269-4. An appropriate I.E.C Approved 10 Amp isolator shall be installed in
both cases.
110 Volt Machine 16 Amp MCB to BS EN60898 Type C or alternatively fused 16 Amp
to I.E.C 60269-4. An appropriate I.E.C Approved 16 Amp isolator shall be installed in
both cases.
Where supplementary protection is provided by an RCD/RCBO, this should be rated at
30 mA.
For all installations, an I.E.E. approved disconnection device must be fitted to
the final installation. This device shall be positioned in close proximity to the
Compact and within easy reach of the operator & shall also be clearly marked as being
the disconnecting device for the machine.