4.9. Display screen icons and warning messages
explanations
Running
End
Blocked Blades
Blocked Drain
Low Water
Power Fail
Hopper Full
Lid cannot open
Lid cannot close
Lid Seal Pressure
Error
The machine has
started and the cycle
has commenced. A
progress bar will be
displayed.
The cycle has ended
and is now ready to
re-use.
The inverter has
detected an overload
due to a blockage of
the cutter blades.
The pressure sensor
has detected a
pressure rise in the
hopper due to a
blocked drain outlet.
There is insufficient
water within the
water tank to allow a
cycle to commence.
An interruption to
the electrical supply
has occurred during
the cycle.
The hopper has not
drained properly, the
outlet may be
blocked.
A fault/obstruction is
preventing the lid
from opening.
A fault/obstruction is
preventing the lid
from closing.
The inflatable seal
has failed to operate
correctly.
Low Deodoriser Lid Lock Fault
Drain Seal Pressure
Error
The drain
(diaphragm) valve
has failed to operate
effectively.
Lid Error
The deodoriser level
is low and should be
refiled at the next
opportunity. The
machine will still run.
The lid interlock has
failed to operate
effectively.
The lid is unable to
open or close.