adopted, which has the advantages of fast refrigeration speed and strong
temperature stability.
2 Control System: Latest proprietary software research and development of
homemade temperature control technology, with high standards of PT100 and all
imports of electronic parts, etc.
3 Circulating System: it is composed of internal and external circulation. During
external circulation, the constant temperature liquid in the tank can be introduced
outside, the second constant temperature field can be established, and it can also
be used as a cold source to lead the cooled liquid in the tank to the experimental
container outside the machine for cooling; During internal circulation, the liquid
temperature in the tank can be very uniform and stable.
4 Protection System: Refrigeration is overheated, over-current protection, call
control system has over temperature alarm, can be set over temperature alarm
temperature, the temperature of the upper and lower limits can cut off the load
automatically when over temperature, and other functions.
5 High-end LCD control instrument, simple operation, temperature stability is
strong, P.i.d can automatically search or artificial adjustment
6 The accurate temperature correction function can reach 0.01℃, and the
maximum temperature fluctuation can reach ± 0.02-0.05℃ (depending on the
model).
Technique Parameter:
Temperature
fluctuation
(±℃)
Operating Steps:
1.Connect the circulating pump before startup (if there is no water inlet and outlet,
this step can be omitted):
1.1 for the connection of circulating pump during internal circulation, connect the
water outlet and water inlet directly with both ends of the hose supplied with the
goods.