- 4 -
1.2 for the connection of circulating pump during external circulation, connect the
water outlet to the inlet of container / equipment outside the tank with hose, and
connect the water inlet to the outlet of container / equipment outside the tank.
2. Add the corresponding liquid medium into the tank. The liquid level of the liquid
medium shall not be 20mm lower than the worktable to avoid damage caused by
exposed dry burning of the heater.
2.1 When the operating temperature is below 8°C, industrial alcohol or antifreeze
is generally used for liquid media.
2.2 At an operating temperature of 8°C-75°C, pure water is generally selected for
liquid media.
2.3 At an operating temperature of 75°C-100°C, dimethyl silicone oil with a
viscosity of 5cs is generally selected for liquid media.
3.Plug in the power, turn on the "main power" switch, and then turn on the "power"
button on the operation panel. (Note: "main power" switch is on the back of the
instrument).
4.Perform the instrument operation as follows:
4.1 Display Description: