Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 40 of 45 DT_0006-6
for 25 cycles
<5 % 100V
(>95 % dip in
100V) for 5 sec
A
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 Α/m
Not
Applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The Model 2120 is intended for use in the electromagnetic
environment specified below. The customer or the user of the Model
2120 should assure that it is used in such an environment
Immunity
test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment -
guidance
Conducted
RF. IEC
61000-4-6
Radiated
RF. IEC
3 Vrms
150 kHz to
80 MHz in
ISM bands
3 V/m
3Vrms from
150kHz top
80kHz
3V/m from
80MHz top
Portable and mobile RF
communications equipment
should be used no closer to
any part of the system,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended
separation distance
d = 1.2
P…80MHz to 800