| Resolution (Flow) | 0.01 L/s |
|---|---|
| Accuracy (Volume) | ±3% or 50 mL, whichever is greater |
| Accuracy (Flow) | Greater of ±5% or 0.200 L/s (BTPS) |
| Connectivity | USB |
| Parameters measured | FVC, FEV1, PEF, FEV1/FVC ratio, FEF25-75%, FET |
| Power Supply | Via USB |
| Resolution (Volume) | 0.01 L |
| Operating Humidity | Up to 95% RH, non-condensing |
| Measurement Range (Volume) | 0 to 10 L |
| Flow range | ±16 L/s |
| Dimensions | 220 x 120 x 60 mm |
| Measurement Range (Flow) | ±16 L/s |
Lists key software features and capabilities for lung function testing.
Guides the initial installation of the Spirotrac software on a PC.
Details configuring a network database connection for Spirotrac.
Explains how to start the software and log in with user credentials.
Covers creating, editing, viewing, and deleting patient data.
Procedures for ensuring device accuracy using a syringe.
Manages user accounts, permissions, and login security.
Covers selecting, backing up, restoring, and merging databases.
General steps and checks before performing pulmonary function tests.
Instructions for performing SpO2 and pulse rate measurements.
Details on various challenge tests like Mannitol and ATS.
Setup and operation of ECG measurements using the BT12 device.
General description and performing hearing threshold tests.
Steps for conducting the 6MWT to assess exercise capacity.
How to record blood pressure and pulse rate measurements.
Essential safety and operational practice using Bacterial Viral Filters.
Guide for diagnosing and resolving software and device problems.
Critical safety information, precautions, and adverse reactions.
Procedures for installing necessary software drivers for connected devices.
Guides on physically connecting various Vitalograph devices to the PC or COMPACT.
Steps to configure wireless network connections for the COMPACT device.