Example –
ADD,DCI,5.0,0.0,0.0,50e-6,,,ABORT
Configures the following (in order) –
A DCI type test step
5 second test time
0 second delay time
No minimum leakage current
50uA maximum leakage current
No arc detection (so following field is also <EMPTY>)
Abort test sequence if fails
PULSE CONFIGURATION FIELDS
POSITIVE - Unipolar +ve
NEGATIVE – Unipolar –ve
BIPOLAR - Bipolar
Ramp time (tr) (in seconds)
Hold Time (th) (in seconds)
Breakdown Current Limit (in A)
Time period for arc detection (4, 10, 15, 20, 30 or 40)
Arc detection limit (in mA)
Only valid if arc detection disabled in field 7
ABORT - Abort on failure
CONT - Continue on failure
Example –
ADD,PULSE,250.0,POSITIVE,1e-3,10e-3,0.05,4,10,ABORT
Configures the following (in order) –
A PULSE type test step
A single positive pulse
1ms ramp time
10ms hold time
50mA maximum current
4us arc detection
10mA arc detection limit
Abort test sequence if fails
BRKDN CONFIGURATION FIELDS