166 2801/2802 Power Analyzer User Guide
1 year Specification, AC Current, Autorange, CF < 2.5, HF Measurements
0.01%
0.10%
1.00%
10.00%
0.0001 0.001 0.01 0.1 1 10 100
Input Level (Arms)
Accuracy (% reading)
Below 100KHz : Add 0.02% per KHz
100-200KHz : Add 0.04% per KHz (tested but
not guaranteed)
Figure 79. AC Current, Autorange, High Frequency