15
Maintenance
There are no field serviceable components inside the Vet-Dop2. Opening the housing
voids the warranty. Store this device in the storage pouch when not in use.
Charging the Battery
The Vet-Dop2 uses a rechargeable lithium-ion battery that has a useful life of
approximately two to four years. When the battery no longer holds a charge, battery
replacement may be required. (NOTE: Call for support concerning battery replacement.)
A full battery charge can take up to 4.5 hours depending upon how depleted the battery
charge is to begin with. The Vet-Dop2 can not be operated when the charger is plugged
in so only charge the battery when the device is not in use.
Specifications
TYPE: CW ultrasonic Doppler blood flow detector
CONSTRUCTION: High impact ABS-PC
DIMENSIONS (OVERALL): 6"x 4.625"x 1.75"
WEIGHT: 15 oz.
DOPPLER FREQUENCY: 9.6 MHz
AUDIO OUTPUT: Speaker and headset jack with speaker cutoff
CHARGER: Input: 110/220 VAC 50 or 60 Hz
Output: 5 VDC (Will not operate unit while charger is connected)
BATTERY: Rechargeable Lithium-ion
BATTERY USE BETWEEN CHARGES: 35-40 hours continuous
TRANSDUCER: Focused piezoelectric elements with 5 Ft. Cable
TRANSDUCER DIMENSIONS: 0.6" X 0.5" X 0.24"
CONNECTOR: 5 Pin circular DIN
SPHYGMOMANOMETER: 0-300 mmHg certified manometer with bulb type
pump and
deflation valve
HEADPHONES: Stereo audio with 3.5 mm phone plug
CUFFS: 6 ea. for limb circumferences 3.3-16.8 cm