42
d) AC current (A ~)
Range Accuracy Resolution
µA
400 µA
±(1.6 % reading +5 digits)
0.1 µA
4000 µA 1 µA
mA
40.00 mA
±(2.0 % reading +8 digits)
10 µA
400.0 mA 0.1 mA
A
4.000 A
±(2.6 % reading +4 digits)
1 mA
10.0 A 10 mA
Overload protection:
mAμA range: F1 fuse 6 x 32 mm F 0.5 A H 600 V (CE) Item no.: 702632
10A range: F2 fuse 6 x 25 mm F 10 A H 600 V (CE) Item no.: 700161
Measuring operations of >5 A may only be performed for at most 10 seconds with a subsequent
measuring pause of 15 minutes.
Frequency response: 45 – 400 Hz
Display: Sine wave valid value (average value response)
e) Resistance
Range Accuracy Resolution
400.0 Ω ±(1.6 % reading +3 digits) 0.1 Ω
4000 Ω
±(1.3 % reading +3 digits)
1 Ω
40.00 kΩ 10 Ω
400.0 kΩ 100 Ω
4.000 MΩ ±(1.5 % reading +10 digits) 1 kΩ
40.00 MΩ ±(2.0 % reading +8 digits) 10 kΩ
Overload protection: 600 V
f) Diode / continuity text
Acoustic continuity tester: ≤10 Ω permanent sound
Acoustic continuity tester resolution: 0.1 Ω
Diode test voltage: 1.5 V
Overload protection: 600 V
g) Non-contact voltage test
Target voltage: 230 V/AC