4. Operation
EzRay Air Portable (Model: VEX-P300) User Manual 39
ENGLISH
System Status depending on Battery Levels
Item
System Status
Battery
Level 3, 2
Battery Level 1 Low Battery
When
turning on
the system
Operation Normal Normal Normal Not operated
Battery
Level
Indicator
Normal Normal Flickers Not displayed
Battery
Charging
Indicator
Not
displayed
Not
displayed
Flickers Not displayed
Control
panel
Brightness
Normal Normal Dark
Normal (Error
code A.10 is
only displayed)
When
operating
the system
Operation Normal Normal Normal Not operated
Battery
Level
Indicator
Normal Normal Flickers Not displayed
Battery
Charging
Indicator
Not
displayed
Not
displayed
Flickers Not displayed
Control
panel
Brightness
Normal Normal Normal
Normal (Error
code A.10 is
only displayed)
Make sure to charge the battery if the Battery Charging
Indicator flickers (or the error code A.10 is displayed). If the
device has been turned on for long periods with the error code
A.10, the battery may be discharged.
DO NOT charge a fully discharged battery, as this may cause
fire or explosion. Be sure to replace the battery (provided by
VATECH).
When the device is connected to the battery charger, the
Battery Charging Indicator is always displayed except that the
battery is fully discharged.