9. Device Specifications
76 EzRay Air Portable (Model: VEX-P300) User Manual
9.3 Electrical Specifications
Item Description
Tube Voltage
Option 1: 60 kV fixed (± 5 %)
Option 2: 65 kV fixed (± 5 %)
Tube Current 2.5 mA (± 10 %)
Exposure Time
0.05-1.0 s (± 3 % or 10 ms)
Rated Voltage
21.6 V
dc
The system will be available with a fixed tube voltage
specification based on the user selection.
9.4 Environmental Specifications
Item Description
During operating
Temperature
10 ~ 35 ℃
Relative humidity 30 ~ 75 %
Atmospheric pressure 860 ~ 1060 hPa
Transport and storage
Temperature
-10 ~ 60 ℃
Relative humidity 10 ~ 75 % non-condensing
Atmospheric pressure 860 ~ 1060 hPa
Failure to follow the specifications above can result in severe
injury and device damage.