EasyManua.ls Logo

VTech EzRay Air Portable - Electromagnetic Compatibility (EMC) Information

VTech EzRay Air Portable
103 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
EzRay Air Portable (Model: VEX-P300) User Manual 95
ENGLISH
A.5 Electromagnetic Compatibility (EMC) Information
Guidance and manufacturers declaration - electromagnetic emissions
The VEX-P300 is intended for use in the electromagnetic environment
specified below. The customer or the user of the VEX-P300 should assure
that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions CISPR
11
Group 1
Class A
The EMISSIONS characteristics of
the VEX-P300 make it suitable for
use in industrial areas and
hospitals (CISPR 11 class A). If
used in a residential environment
(for which CISPR 11 class B is
normally required), this equipment
might not offer adequate
protection to radiofrequency
communication services. The user
might need to take mitigation
measures, such as relocating or
re-orienting the equipment.
Harmonic emissions
IEC 61000-3-2
Applicable
The VEX- P300 is suitable for use
in all establishments and may be
used in domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Applicable