Appendix
EzRay Air Portable (Model: VEX-P300) User Manual 95
ENGLISH
A.5 Electromagnetic Compatibility (EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions
The VEX-P300 is intended for use in the electromagnetic environment
specified below. The customer or the user of the VEX-P300 should assure
that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions CISPR
11
Group 1
Class A
The EMISSIONS characteristics of
the VEX-P300 make it suitable for
use in industrial areas and
hospitals (CISPR 11 class A). If
used in a residential environment
(for which CISPR 11 class B is
normally required), this equipment
might not offer adequate
protection to radiofrequency
communication services. The user
might need to take mitigation
measures, such as relocating or
re-orienting the equipment.
Harmonic emissions
IEC 61000-3-2
Applicable
The VEX- P300 is suitable for use
in all establishments and may be
used in domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Applicable