EasyManua.ls Logo

VTech EzRay Air Portable - X-Ray Dose Data

VTech EzRay Air Portable
103 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
86 EzRay Air Portable (Model: VEX-P300) User Manual
A.4 X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the device. The
IEC collateral standards measured the X-ray doses of the device in the test report. The
device was designed by Part 1. General Requirements for Safety, IEC 60601-1-3.
Test Condition
Model Name VEX-P300
Tube Model Name V1-650304
Generator Model Name DG-S0101V1 (Inverter type)
Loading Factor
Option 1: 60 kV, 2.5 mA
Option 2: 65 kV, 2.5 mA
A.4.1 X-ray Dose Table
Test Equipment
Instrument Manufacturer Model S/N
Multi-Dose Meter Raysafe
Unfors Xi mAs
/ Unfors Xi R/F & MAM
163288 / 161834
Digital Sensor
Dose Area Product (DAP) Table
(60 kVp, 2.5 mA, SSD 200 mm)
FOV: Ø 6 cm FOV : 3 x 4 cm FOV : 2 x 3 cm
t (s) Dose (mGy.cm
2
)
0.14 6.5 2.8 1.4
0.16 7.5 3.2 1.6
0.18 8.5 3.6 1.8
0.20 9.8 4.2 2.1
Dose Area Product (DAP) Table
(65 kVp, 2.5 mA, SSD 200 mm)
FOV: Ø 6 cm FOV : 3 x 4 cm FOV : 2 x 3 cm
t (s) Dose (mGy.cm
2
)
0.12 6.7 2.9 1.4
0.14 7.9 3.4 1.7
0.16 9.1 3.9 1.9
0.18 10.2 4.3 2.2