Appendix
EzRay Air Portable (Model: VEX-P300) User Manual 77
ENGLISH
Appendix
A.1 How to Attach a Device to a Tripod
Some countries do not allow using a hand-held X-ray in clinics. Users in those
countries must use a Tripod and Remote Exposure Switch. Read and follow the
instruction below to mount the device to the tripod and connect the Remote Exposure
Switch using a Base Holder provided by VATECH.
Base holder, Remote Exposure Switch, Cables, and tripod
are provided as options and shipped in a separate package.
See ‘Available Option Items’ in Chapter 3.
Demount the device from the tripod by following each step in
reverse order.
WorkForce
1 or 2 person(s)
Required tool
A screwdriver with a magnetic tip (preferred)
Mounting Procedure
1. Prepare the Base Holder, Tripod, and Remote Exposure Switch with a cable. The base
holder’s bottom and upper parts are connected with 4 cross-head screws. (See the
below)