7. Cleaning and Maintenance
EzRay Air Portable (Model: VEX-P300) User Manual 67
ENGLISH
7.2.1 Maintenance Task Checklist
Always turn off the device before performing any maintenance.
Tasks Period
Before the operation, ensure that the device is clean and ready
for use.
Daily
After using the device, make sure that the device has been
turned off.
Daily
Wipe the device's outer covers with a dry cloth at the end of
each day’s operation.
DO NOT use detergents or solvents to
clean the outer covers of the device.
Daily
Ensure that the signal is audible and the X-ray emission light is
visible when making an exposure.
Daily
Ensure that the yellow (exposure) indicator light turns on when
the Exposure Button is pressed.
Daily
Ensure that the battery charging LED indicator comes on when
charging the battery.
Daily
Ensure that the battery level indicator displays at least two levels
(Battery Level 2). For more information on the battery levels, see
4.6 Using the Battery’.
Daily
Ensure that all visible labels are intact and legible. Monthly
If any defects are found, do not operate the device since the
problem must be handled only by qualified personnel.