Appendix
90 EzRay Air Portable (Model: VEX-P300) User Manual
A.4.3 Scattered Dose
Scope
IEC 60601-2-65 203.13
Requirements
ME EQUIPMENT shall be provided with means to optionally allow actuation of the
EXPOSURE from a PROTECTED AREA after installation.
Relevant instructions shall be given in the ACCOMPANYING DOCUMENTS.
Results
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively.
Test Equipment
Instrument Manufacturer Model S/N
X / Gamma
Survey Meter
Radcal Co. 9015/10X5-180 91-1470/19069
Method
PMMA Phantom aligned to 280 mm away from Focal Spot
Max. Exposure Condition
Measure point: 1000 mm from PMMA Phantom