EasyManua.ls Logo

WarmTouch WT-5900 - Temperature Control

WarmTouch WT-5900
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the Warming System
22 Operator’s Manual
Figure 8. Low Temperature Light
Temperature Control
WARNING
The patient must be closely monitored for rewarming. Vasodilation and
possible hypotension can occur. Use good judgment when selecting a
temperature. If unsure of proper setting, consult with the attending
physician.
WARNING
Using the warming system on transdermal medication patches may increase
the rate of drug delivery, potentially causing harm to the patient.
The control panel features a warning light indicator and four manually-switched
temperature settings. The warning light indicates the control system has
identified an alarm condition and an action by the operator is required. Each
temperature setting represents the average temperature of air entering the
hose.
Low 32°C (89.6°F)
Medium 38°C (100.4°F)
High 43°C (109.4°F)
Boost 45°C (113°F) for 45 minutes
For patients with severe hypothermia requiring rapid warming, select the Boost
temperature setting. To help prevent hypothermia or for treatment of mild
hypothermia, select High, Medium or Low temperature settings. The selected
temperature light illuminates.