Overview
Operator’s Manual 27
1 Specifications
Overview
This chapter contains physical and operational specifications for the
WarmTouch™ Model WT-5900 patient warming system.
Warming System Specifications
Table3.System Specifications
Warming Blanket Specifications
Maximum blanket surface temperature 44
°C
Blower Specifications
Dimensions 38 cm x 41 cm x 28 cm
(15 inches x 16 inches x 11 inches)
Weight 6.8 kgs (15 lbs.)
Power Requirements 220-230 volts AC, 50/60 Hz, 6 amp
Automatic Temperature Stepdown
(Boost to High Temperature)
After 45 minutes of continuous use blower will
step down from the Boost to High setting.
Power Supply Cord 4.26 m (14 feet)
Thermal Protection Threshold Thermostat (internal): 47
°C to 50°C
(117
°F - 122°F)
Ambient Blower
Operating Temperature Range
18
°C - 28°C (64.4°F - 82.4°F)
Over Temperature Alarm Level 65 dB at 3 meters
Protection Against Ingress of Fluids Ordinary
Cart Specifications
Weight 3.1 kg (6.8 pounds)
Height 67.1 cm (26.4 inches)
Width 32.3 cm (12.7 inches)
Depth 38.6 cm (15.2 inches)