No communications with the XEVO TQD, TQ-XS, TQ-S micro & SQD2 -
WKB7610
SYMPTOMS
Unable to control MS in MassLynx tune page
Loss of comms with MS
"Instrument Not Present" error message in MassLynx pane (bottom right)
ENVIRONMENT
MassLynx 4.1
TQD
Xevo TQD
Xevo TQ-XS
Xevo TQ-S micro
SQD2
CAUSE
Embedded PC does not have communications with the host PC.
FIX or WORKAROUND
1. Close MassLynx and the MS console.
2. Locate the EPC reboot switch on the front of the instrument (small hole just below the power and status LED) or
for TQD check inside front panel door next to electrical and gas connections .
3. Push a piece of peek tubing or a paper clip into the small hole mentioned in step 2. This is where the EPC reboot
switch is located.
4. Wait three minutes for the EPC to run through its boot sequence.
5. Open MassLynx and confirm that a "Not Scanning" message is in the MassLynx pane (bottom right).
6. Open MS Tune page and confirm vacuum readings are "real" and that everything is communicating properly. Note:
If the MS has vented, go to Vacuum > Pump to start the backing pump.
7. Within the MS tune page, confirm that MS API gas can be turned on, the MS into Operate mode and that there are
sensible readbacks.
Waters - The Science of What's Possible.
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB7610_No_communications_with_the_XEVO_TQD
Updated: Fri, 17 Sep 2021 00:04:37 GMT
Powered by
1

Other manuals for Waters Xevo TQ-S micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters Xevo TQ-S micro and is the answer not in the manual?

Waters Xevo TQ-S micro Specifications

General IconGeneral
BrandWaters
ModelXevo TQ-S micro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals