Hygienic preparation 25
EN
5. Hygienic preparation
This product may contain disposable items. Disposable items are intended to be
used only once. So use these items only once and do
not reprocess them. Repro-
cessing disposable items may impair the functionality and safety of the product
and lead to unforeseeable reactions as a result of ageing, embrittlement, wear,
thermal load, the effects of chemical processes, etc.
5.1 Intervals
Clean the mask and exhalation system every day in accordance
with the operating instructions.
Check the filters regularly:
clean the coarse dust filter once a week and replace it at least every 6
months.
change the fine filter once a month depending on soiling.
Regularly wipe clean the housing of the CPAP 20e and the power
supply.
The breathing hose should be cleaned according to the level of
soiling, however at least once a month. Replace it every 12
months.
Wash the headgear in accordance with the accompanying in-
structions for usage.
Clean the humidifier every day.