34 Functional check
EN
6.3 Function check of CLICK 2
1. Carry out a visual inspection of the plastic housing.
If cracks/damage or severe soiling is evident, the plastic parts or seals must be
replaced.
2. Fill the humidifier up to the mark with water.
3. Check whether the humidifier is leaking.
4. Empty the water.
5. Now pour in about 50 ml of water.
6. Connect the humidifier to the therapy unit.
7. Switch on the therapy unit.
8. Switch the humidifier on by pressing the humidifier button on the therapy
unit.
9. Set heater level 6 on the therapy unit.
10.Check whether the humidifier is getting warmer.
If the water has not heated up after 15 minutes, send the humidifier and the
therapy unit for repair to the dealer or the manufacturer.