EasyManua.ls Logo

Weinmann CPAP 20e - Disinfection of CLICK 2

Weinmann CPAP 20e
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hygienic preparation 31
EN
We recommend that you wear suitable gloves when disinfecting the equipment
(e.g. household or disposable gloves).
Breathing hose
We recommend gigasept FF
®
as a disinfectant.
When disinfecting use the same procedure as for cleaning.
The breathing hose must not be heated to temperatures in excess
of 70 °C. Sterilization is not permitted.
Accessories
To disinfect or sterilize the exhalation systems and the mask, please see the sec-
tion "Hygienic preparation" in the relevant operating instructions.
When there is a change of patient please refer to section “5.6 Disinfection
CLICK 2” on page 31 and “5.8 Multiple patient use CLICK 2” on page 32 for in-
structions on how to disinfect the CLICK 2 humidifier.
5.6 Disinfection CLICK 2
The humidifier can be disinfected. Please observe the operating instructions for
the disinfectant used. We recommend that you wear suitable gloves when disin-
fecting the equipment (e.g. household or disposable gloves).
We recommend gigasept FF
®
.
The procedure for dismantling and assembling the humidifier is
the same as for cleaning.
To ensure full disinfection, the filling cap should also be removed
and disinfected separately.
For hygiene reasons, we recommend that the plastic parts are replaced after a
maximum of two years use.