Technical data 43
EN
Subject to design modifications.
All values established subject to ATPD conditions (ambient temperature and
pressure, dry).
CLICK 2
Product class as per 93/42/EEC II a
Dimensions WxHxD in mm 120 x 97 x 157
Weight (without water) 250 g
Temperature range
operation
storage
+5 °C to +35 °C
–20 °C to +70 °C
Humid operation and storage 95 % relative humidity
Ambient pressure range 600 to 1100 hPa
Electrical connection 30 V DC
Power consumption 20 VA
Classification as per EN 60601-1
level of protection from electric shock Unit type B
Electromagnetic compatibility (EMC) as per
EN 60601-1-2
Radio shielding
radio interference suppression
EN 55011 B
IEC 1000-4, parts 2 to 6, part 11
Max. permitted fill quantity 280 ml
Max. permitted operating pressure 30 hPa
Max. permitted flow (free-flowing) 160 l/min
Max. mask temperature 37 °C
Gas leakage at 30 hPa negligible
Humidifier output at heating level 6:
Flowrate = 20 l/min
Flowrate = 30 l/min
Flowrate = 40 l/min
at 23 °C and 65 % relative humidity
14.1 mg/l
10.3 mg/l
8.0 mg/l