Safety
Hazardous therapy due to leaks during
ventilation!
The measured MVi value indicates the volume the device is administering
to the patient. In the event of leaks during ventilation, the measured MVi
value will not match the tidal volume actually administered. If this is not
observed, the patient might suffer serious or life-threatening injury.
à Do not use the measured MVi value as an adequate means of
assessing ventilation.
à Use external monitoring (etCO
2
or expiratory volume measurement).