Ventilation device
Description of device and instructions for use
for devices from serial number 20,000
VENTIlogic LS
VENTIlogic plus

Table of Contents

Questions and Answers

  • Z
    Zachary ArroyoSep 15, 2025
    What to do if the Weinmann VENTIlogic plus Medical Equipment displays 'Internal battery not detected'?
    • R
      Russell LewisSep 16, 2025
      If your Weinmann Medical Equipment displays 'Internal battery not detected', it may be due to a defective battery or the use of a non-approved battery. Have the device repaired.
  • S
    Stacie FloydAug 25, 2025
    What to do if Weinmann VENTIlogic plus Medical Equipment shows O 2low?
    • E
      elizabeth19Aug 25, 2025
      If the Weinmann Medical Equipment displays an 'O 2low' message, it could be due to several reasons: the oxygen flow rate might be set too low, there could be a leak in the system, the oxygen supply might be interrupted, or the oxygen sensor may be incorrectly calibrated. Check the oxygen flow rate at the source, look for and eliminate any leaks, ensure the oxygen supply and connections are secure, and calibrate the oxygen sensor.
  • C
    Christina OlsonAug 26, 2025
    Why is Weinmann VENTIlogic plus Medical Equipment showing O 2high?
    • A
      Anthony FrenchAug 26, 2025
      If your Weinmann Medical Equipment is showing 'O 2high', this could be due to the oxygen flow rate being set too high or the oxygen sensor being incorrectly calibrated. Verify the oxygen flow rate matches the doctor's prescription at the oxygen source. If necessary, have the settings checked by the doctor. You may also need to calibrate the oxygen sensor.
  • E
    Eric AndersonAug 26, 2025
    What to do if Weinmann Medical Equipment internal battery is exhausted?
    • J
      Jonathan RossAug 27, 2025
      If the internal battery of your Weinmann Medical Equipment is exhausted, press the alarm acknowledgement key. Then, have the battery replaced by a specialist dealer to ensure the therapy data is correctly recorded.
  • C
    Cynthia SmithSep 16, 2025
    Why won't my Weinmann VENTIlogic plus Medical Equipment internal battery charge?
    • J
      james37Sep 16, 2025
      The internal battery of your Weinmann Medical Equipment may not be charging because it is being operated outside the permitted temperature range. Protect the device from direct sunlight and do not operate it close to a radiator. Ensure that the device is operated within the permitted temperature range.
  • D
    Dennis BrockSep 23, 2025
    Why softstart time cannot be set to 30 min. on Weinmann VENTIlogic plus?
    • A
      Angela GordonSep 23, 2025
      The maximum softstart time is limited by your doctor. Clarify with your doctor whether the maximum softstart time for your therapy can be extended to 30 minutes.
  • B
    bradley26Sep 8, 2025
    What does BT low mean on Weinmann VENTIlogic plus Medical Equipment?
    • S
      scottjonesSep 8, 2025
      If the Weinmann Medical Equipment displays 'BT low', the settings may be implausible. Have the settings checked by the doctor supervising treatment.
  • C
    Christopher McclureSep 9, 2025
    What to do if Weinmann VENTIlogic plus Medical Equipment displays service life ended for replaceable battery?
    • M
      mmaddenSep 9, 2025
      If your Weinmann Medical Equipment displays 'Service life ended. Have replaceable battery replaced', it means the replaceable battery has reached the end of its service life. Replace the replaceable battery.
  • R
    Richard PetersonSep 2, 2025
    What to do if Weinmann VENTIlogic plus Medical Equipment doesn't detect replaceable battery?
    • Z
      zwilliamsSep 2, 2025
      If your Weinmann Medical Equipment is not detecting the replaceable battery, it could be due to several reasons: the battery may be defective, a non-approved battery is in use, or you are using the WM 27998 battery with a firmware version below 2.9.0. Have the device repaired if the battery is defective. Use a genuine Weinmann battery. Update the firmware to version 2.9.0 or above if using the WM 27998 battery.
  • J
    Jose JimenezAug 31, 2025
    Why is the SpO signal weak on Weinmann Medical Equipment?
    • S
      Stephen JonesSep 1, 2025
      If the SpO signal is weak on your Weinmann Medical Equipment, check the SpO sensor and fingertip and clean if necessary.

Summary

2. Description of device

Intended Use of the Device

Details the applications and patient suitability for VENTIlogic LS and VENTIlogic plus devices.

Description of Device Function

Explains how the device operates, including pressure delivery, display, and modes.

3. Safety instructions

General Safety Information

General safety advice, warnings, and cautions for using the device safely.

Life-Support Ventilation Safety

Critical safety warnings regarding resistance and potential alarm failures in life-support modes.

Safety Guidelines for Device Operation

Safety guidelines for device operation, environment, and patient interface use.

Contraindications for Device Use

Lists diseases or conditions where the device should not be used or used with caution.

4. Set up device

Setting Up and Connecting the Device

Instructions for physically setting up the device and making initial connections.

Connecting Valve Ventilation

Steps for connecting the patient circuit for valve ventilation modes.

Connecting Leakage Ventilation

Instructions for connecting the patient circuit for leakage ventilation.

Connecting the Bacteria Filter

Instructions for connecting bacteria filters to prevent contamination.

Therapy with Oxygen Supply

Procedures for supplying and monitoring oxygen therapy with the device.

Operation During Power Failure

How the device operates and what to do during a power outage.

5. Operation

Device Controls Overview

Overview of the device's controls and their functions.

Starting Up the Device

Steps for starting up the therapy device and its operating states.

Handling Device Batteries

Information on internal and replaceable batteries, charging, and status.

Accessing the Alarm List

How to access and understand the stored alarm history.

Selecting a Therapy Program

How to choose and set pre-configured therapy programs.

Post-Operation Procedures

Steps to take after therapy, including device shutdown and cleaning.

6. Hygiene treatment

Cleaning for Leakage Ventilation

Detailed steps for cleaning patient circuits used in leakage ventilation.

Cleaning the Device Housing

Instructions for cleaning the exterior of the therapy device safely.

Coarse Dust Filter Cleaning and Fine Filter Change

Procedure for cleaning the coarse dust filter and changing the fine filter.

Disinfection and Sterilization Guidelines

Guidelines for disinfecting device components and accessories.

7. Function check

Function Check Method

General method for performing function checks on various device modes.

Check Flow;Pressure Sensors (Leakage Ventilation)

Procedure to check flow and pressure sensors in leakage ventilation mode.

Calibrating the Oxygen Sensor

Step-by-step guide for calibrating the oxygen sensor.

8. Troubleshooting

Device Faults and Remedies

Table of common device faults, their causes, and recommended remedies.

9. Maintenance and safety checks