EasyManuals Logo

Weinmann VENTIlogic plus User Manual

Default Icon
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Maintenance and safety checks 95
EN
9. Maintenance and safety checks
9.1 Intervals
We recommend having servicing, safety checks and repair work carried out only by
the manufacturer, Weinmann, or by a specialized dealer expressly authorized by
the manufacturer.
Check both filters regularly for dirt.
The coarse dust filter and the fan filer should be cleaned once a week and
changed no later than every 6 months.
The fine filter must be changed after no more than 1000 operating hours (filter
change symbol appears in the display).
For hygiene reasons, we recommend replacing the following parts at the intervals
specified.
9.1.1 Leakage ventilation
Pressure measuring tube every 6 months - earlier if dirty
Clean the patient/ventilator interface in accordance with the relevant
instructions for use
Clean the exhalation system in accordance with the relevant instructions for use
If necessary, other accessories in accordance with the associated operating
instructions
9.1.2 Valve ventilation
The patient circuits with patient valve should be replaced in accordance with the
relevant instructions for use
Clean the patient/ventilator interface in accordance with the associated
instructions for use
If necessary, other accessories in accordance with the associated instructions for
use

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic plus and is the answer not in the manual?

Weinmann VENTIlogic plus Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals