11.2 System resistances
System resistances do not change in the first instance of a fault.
11.3 Bacteria filter WM 24148 and WM 27591
System resistance at an air flow rate of 60 l/min at the patient connection opening
Accessories
Therapy device with
tube system
WM 24130 (leakage
system) and
Silentflow
WM 23600
Therapy device
with single patient
circuit with patient
valve
VENTIlogic LS with
double patient
circuit with patient
valve
Standard configuration
0.24
0.38 0.4
O
2
sensor WM 27128
-
0.47 0.49
Bacteria filter
WM 27591
-
0.48 0.5
O
2
sensor WM 27128
and bacteria filter
WM 27591
-
0.57 0.59
Bacteria filter WM 24148 for leakage ventilation
Product class to Directive 93/42/EEC IIa
Dimensions Ø x L in cm 7.4 x 9.8
Weight approx. 51 g
Temperature range
– operation
– storage
+5 °C to +40 °C
-20 °C to +70 °C
Permitted humidity in operation and storage 95 % rh (no
condensation)
Ambient pressure range 700 to 1060 mbar