124 Technical data
EN
11.12 Electromagnetic interference immunity
for NON-LIFE-SUPPORTING medical
electrical devices or medical electrical
systems
Guidelines and manufacturer declaration - electromagnetic INTERFERENCE
IMMUNITY
The VENTIlogic LS/VENTIlogic plus is intended for operation in the ELECTROMAGNETIC
ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/
VENTIlogic plus should ensure that it is used in an environment of this type.
INTERFERENCE
IMMUNITY tests
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVI-
RONMENT guideline
Portable and mobile radio
devices should not be used
at a distance
from the VENTIlogic LS
(including its cables) of less
than the recommended
safety distance calculated in
accordance with the
equation applicable to the
transmission frequency.
Recommended safety
distance:
Conducted HF
interference to IEC
61000-4:-6
3 V
effective value
150 kHz bis 80
MHz
10 V 1.7 m
Radiated HF
interference to IEC
61000-4:-3
3 V/m
80 MHz bis 2.5
GHz
10 V/m
1.7 m for 80 MHz to 800
MHz
1 m for 800 MHz to 2.5 GHz