Hygiene treatment 71
EN
6.5 Clean the fan filter
The fan filter protects the housing fan from dirt.
To clean the fan filter, proceed as follows:
1. Remove the fan filter according to the instructions in section “Change fan filter” on
page 99.
2. Clean the fan filter with fresh running water until it is free of residue.
3. Let the fan filter dry. The fan filter must be completely dry before starting the device.
4. Insert the fan filter again according to the instructions in section “Change fan filter”
on page 99.
6.6 Clean the accessories
To clean the accessories, see the section entitled "Hygiene treatment" in the
corresponding instructions for use.
6.7 Clean the SpO
2
module
The housing of the SpO
2
module should be cleaned at regular intervals depending on
contamination.
Wipe the SpO
2
module and the connecting cable with a soft damp cloth.
6.8 Disinfect, sterilize
If required, e.g. following infectious diseases or unusual contamination, you can also
disinfect the housing, the power cord, the patient circuit (leakage ventilation only) and the
bacteria filter housing (leakage ventilation only). See the instructions for use for the
disinfectant used. We recommend wearing suitable gloves (e.g. household or disposable
gloves) for disinfecting.
6.8.1 Device
The housing and the power cord of the therapy device are cleaned simply by wiping with
disinfectant. We recommend terralin
®
protect for this purpose.