EasyManua.ls Logo

Weinmann VENTIlogic plus - Check Flow;Pressure Sensors (Valve Ventilation); Checking Device Alarms

Default Icon
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 Function check
EN
7.2.2 Check flow measurement and flow sensors/pressure sensors
(leakage ventilation with leakage adapter for standard
tapered connector)
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. Perform the subsequent steps as descibed in “7.2.1
Check flow measurement and flow sensors/pressure
sensors (leakage ventilation)” from step 2.
7.2.3 Check flow measurement and flow sensors/pressure sensors
(valve ventilation)
Function check using a single patient circuit with patient valve
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. The subsequent steps are described in “7.2.1 Check
flow measurement and flow sensors/pressure sensors
(leakage ventilation)” starting from step 2.
Function check using a double patient circuit with patient valve (VENTIlogic LS
only)
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. The subsequent steps are described in “7.2.1 Check
flow measurement and flow sensors/pressure sensors
(leakage ventilation)” starting from step 2.
7.2.4 Alarms
The therapy device performs a self-test on the sensor system when the On/Off key is
pressed which also checks that the alarm system is working. If a fault occurs during the self-
test, an error message appears in the main screen (see also “8. Troubleshooting” on
page 81).
1. Check buzzer and LEDs:

Table of Contents

Related product manuals