Safety
Disrupted or failed therapy due to incorrect use
of disposables!
Reusing and reprocessing disposables might induce unpredictable
reactions as a result of aging, embrittlement, wear, thermal loading and
chemical action. This might place the functionality and safety of the
device at risk, and cause the patient or user to suffer serious or life-
threatening injury.
à Do not reuse disposables.
à Do not subject disposables to hygienic reprocessing.
Risk of injury resulting from condensation in the patient valve at
temperatures below 0 °C!
If patients are subjected to sustained ventilation, at temperatures
below 0 °C, the exhalation moisture in the patient valve might
condense and impair the function of the parts. This might expose the
patient to the risk of serious or life-life-threatening injury.
à Quickly move the patient to a warmer location.
à At temperatures below 5 °C, use a breathing system filter to extend
application time.
SKR 11/2019 MEDUVENT Standard User Training © WEINMANN