CP 50 and CP 50 Plus 12-lead resting electrocardiographs — software update 01.30.04 5
Update the software
1. Make sure the system is powered down. To power down the CP 50, press and hold the power button
for 6 seconds.
2. Connect the CP 50 to AC power.
3. Plug the USB drive (part number 411053) into the CP 50.
4. Power up the CP 50 by pressing the power button. The software update begins after the device
powers up.
5. When the software update completes, the device might prompt you for the serial number and the
manufacturing date. Enter these values and click Update.
6. Remove the USB drive after the update is complete. Press OK to restart.
7. Once the device has completed the restart process, select your language.
Verify the software version:
a. Go to Menu > Settings (System tab) > About.
b. Verify that the correct software version is listed for the CP 50.
Note
The device will restart multiple times if the current software version 01.20.03 or later. The
software update will operate on all earlier versions.
Note
The serial number is located on the label in the rear of the device. The current date can be
entered of the manufacturing date is unavailable.
Caution Do NOT remove AC power from the device during the restart.

Other manuals for Welch Allyn CP 50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn CP 50 and is the answer not in the manual?

Welch Allyn CP 50 Specifications

General IconGeneral
BrandWelch Allyn
ModelCP 50
CategoryMedical Equipment
LanguageEnglish

Related product manuals