ELI 150
5-4
Patient Input
Connector:
Pin #:
Name:
Description:
1
V2
Patient Lead
2
V3
Patient Lead
3
V4
Patient Lead
4
V5
Patient Lead
5
V6
Patient Lead
6
SHIELD
Isolated Front-End Ground
7
Reserved
8
Reserved
9
RA
Patient Lead
10
LA
Patient Lead
11
LL
Patient Lead
12
V1
Patient Lead
13
Reserved
14
RL
Patient Lead
15
Reserved