Faults Causes Measures
No pointer movement
despite change in
pressure.
Movement blocked. Replace instrument.
Pressure element defec-
tive.
Pressure port blocked.
After depressurisation,
the pointer remains just
above the zero point.
 Tap lightly on the
case.
Instrument was overloaded. Replace instrument.
Material fatigue of the
pressure element.
The pointer remains
outside the zero
point tolerance after
installation and
depressurisation.
Mounting error: Instrument
not mounted in nominal
position.
Check the mounting
position.
Transport damage (e.g.
non-permissible shock
loading).
Replace instrument.
Instrument outside the
accuracy class.
Instrument was operated
outside of permissible
performance limits.
Check the obser-
vance of the operat-
ing parameters of the
application.
Replace instrument.
Vibration of the pointer. Vibrations in the appli-
cation.
Use instrument with

Mechanical damage
(e.g. window, case).
Improper handling. Replace instrument.
Impermissible loading at
the installation point (e.g.

Security tag damaged
(models PG21BP-x)
Manipulation of the cali-
bration possible.
Send the instrument
for recalibration.