36 WIKA operating instructions, model CPG1200
EN
14603921.02 06/2023 EN/DE
7.2.5 Damping
The filter influences the currently displayed measured value.
The measured value of the sensor, which is used for other functions, is not affected.
The following factors can be selected:
Off
Medium = 0.8
Low = 0.6
High = 0.9
The calculation will be carried out as per the following formula:
Indication value = last indication value * factor + current measured value * (1-factor)
Tare is a temporary offset and has no effect on the measurement
performance of the sensor.
To change, activate or deactivate the damping, proceed as follows under “Menu” /
“Measure mode” / “Damping”:
1. With the [LIGHT/▼] button or [ZERO/▲] button, select the desired function.
2. Confirm the setting by pressing the [MENU/◀▶] button.
The instrument moves back one menu level.
Press the [HOME] button in order to exit the menu.
7.2.6 Sample rate
The sample rate indicates the interval at which the CPG1200 requests pressure values
from the sensor.
The following values can be selected:
1/s
2/s (default)
4/s
10/s
Logger interval
To change the sample rate, proceed as follows under “Menu” / “Measure mode” /
“Sample rate”:
1. With the [LIGHT/▼] button or [ZERO/▲] button, select the desired rate.
2. Confirm the setting by pressing the [MENU/◀▶] button.
The instrument moves back one menu level.
Press the [HOME] button in order to exit the menu.
If the “Logger” / “Interval” function is selected, measuring rate = logger interval is
automatically set.