EasyManua.ls Logo

WIKA RTD - 9.1 RTD Measuring Point with Thermowell Example

WIKA RTD
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WIKA operating instructions RTD and TC, intrinsically safe designs
3345267.08 06/2011 GB/D
33
GB
9.1 Calculation example for RTD measuring point with thermowell
Use at the partition wall to Zone 0: Calculate the maximum permissible temperature T
max
at the
thermowell tip for the following combination:
RTD measuring insert Ø 6 mm with built-in model T32.1S head-mounted transmitter, tted into
a Design 3F multi-part thermowell . Power supply is, for example, via a Model KFD2-STC4-EX1
transducer power supply (WIKA Article No. 2341268).
T
max
is obtained by adding the temperature of the medium and the self-heating. The self-heating
of the thermowell tip depends on the supplied power P
o
of the transmitter and the thermal
resistance R
th
.
The following formula is used for the calculation: T
max
= P
o
x R
th
+ T
M
T
max
= Surface temperature (max. temperature at the thermowell tip)
P
o
= from transmitter data sheet
R
th
= Thermal resistance [K/W]
T
M
= Temperature of the medium
Prerequisite is an ambient temperature T
amb
of -20 ... +40 °C.
Example
Resistance thermometer RTD
Diameter: 6 mm
Temperature of the medium T
M
= 150 °C
Supplied power: P
o
= 15.2 mW
Temperature Class T3 (200 °C) must not be exceeded
Thermal resistance [R
th
in K/W] from table = 37 K/W
Self-heating: 0.0152 W x 37 K/W = 0.56 K
T
max
= T
M
+ self-heating: 150 °C + 0.56 °C = 150.56 °C
The result shows that in this case self-heating at the thermowell tip is negligible. As safety
clearance for type-examined instruments (for T6 to T3), another 5 °C must be subtracted from the
200 °C; hence 195 °C would be permissible. This means that in this case temperature class T3 is
not exceeded.
Additional information
Temperature class for T3 = 200 °C
Safety clearance for type-examined instruments (T6 to T3)
*1
= 5 K
Safety clearance for type-examined instruments (T1 to T2)
*1
= 10 K
*1 EN/IEC 60 079-0: 2009 Section 26.5.1
9. Calculation examples for self-heating at the sensor/thermowell tip

Related product manuals