GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The Willow Go™ Breast Pump is intended for use in the electromagnetic environment specied below. The customer or user of the Willow Go™
Breast Pump should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment – Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
Normal operation while
on battery power and
Battery Charge Mode
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, and
±15 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, and
±15 kV
Floors should be wood, concrete, or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply lines ±±2 kV for power supply lines
Mains power quality should be that of
a typical commercial, home or hospital
environment.
Surge
IEC 61000-4-5
± 0.5 kV, ±1 kV line to line ± 0.5 kV, ±1 kV line to line
Mains power quality should be that of a typical
commercial, home or hospital environment.