EasyManua.ls Logo

wtw 315i - Configuring; Display During Calibration Setting; Configuration Procedure

wtw 315i
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring pH 315i
8
Configuring
The following setting can be adapted individually to your requirements (the
delivery status is shown in bold typeface):
Procedure for configuring (factory settings appear in bold typeface):
Note
At this point, the AutoCal TEC calibration can be terminated with
<M>. This corresponds to a two-point calibration. The instru-
ment uses the slope and the asymmetry of calibration range I.
l Immerse the pH electrode in the second buffer solution to continue the
two-point calibration.
l Press <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) or the buffer nominal value appears on the display.
l If the measured value is stable, the instrument displays the value of the
slope and the evaluation of the calibration in calibration range II.
l Press <RUN/ENTER>. The instrument displays the value of the asymme-
try in calibration range II. The three-point calibration is completed.
l Switch to the measuring mode with <M>.
8
4
2
3C
t
°C
TP
pH
AutoCal TEC
pH
85
8
mV/pH
-
pH
0
mV
Display during pH calibration Buffer nominal value, or current
electrode voltage
l Switch off the measuring instrument.
l While pressing the <M> key, press <>.
l Select the required display during the pH calibration with <CAL>.
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l Then press the <RUN/ENTER> key. The measuring instrument automat-
ically switches to the measuring mode.
LC
A
pi
sd
/pH

Related product manuals