XOS | 801377-02 | Rev D | English Page 4 of 22
System Specifications –
Electrical, Physical, and Operational
PetraMAX™ - Range, LOD & Applications
Dynamic Range Sulfur 5.7 ppm - 10 wt.%
Limit(s) of
Detection
(ppm @ 600 s)
Sulfur 5.7 ppm
P Cl K Ca V Cr
17 3 0.7 0.4 0.1 0.09
Mn Fe Co Ni Cu Zn
0.07 0.07 0.07 0.04 0.1 0.1
Applications
Elements in liquid hydrocarbons like
crude oil, diesel, gasoline, jet fuel, and
lubricants. Also elements in catalyst
powder, petroleum coke / coal and
Petra4294™ - Range, LOD & Applications
Dynamic Range Sulfur 2.6 ppm - 10 wt.%
Limit of Detection Sulfur 2.6 ppm at 600s
Applications
Sulfur in hydrocarbons like crude oil,
diesel, gasoline, jet fuel, and lubricants
Method Compliance
Measurement Time 30 - 900s
Calibration 30 calibration curves
Sample Film
PetraMax: Etnom®, 3.0 µm
Petra4294: Mylar, 3.6 µm
Sample Fill Volume 8 mL minimum
Data Output
Printout ● USB
Ethernet to PC connection
I/O Ports Ethernet 10/100, USB
Power Supply 110-240 VAC ± 10%, 50-60 Hz
Operating Temperature 41°F to 104°F (5°C to 40°C)
Operating Humidity 30 – 85 %
Weight 28 lbs. (12.7 kg)
Dimensions (W x D x H)
14.5 x 16.5 x 6.0 in.
36.9 x 41.9 x 15.2 cm
Etnom® is a registered trademark of ChemPlex Industries, Inc.