Chapter 5 Error message
Error message and measures for mis-operation
1. Err 1 AD value is too small when calibrating (please select the load cell with
proper capacity)
2. Err 2 Inputted weight data is 0 when full load calibrating in calibrating mode.
(Please input the correct weight data of the weights on the platform)
3. Err 3 Zero position exceeds the setting range when turning the indicator on.
(Please ensure the weight on the scale is 0 when turning the indicator on)
4. Err 4 Single weight is below 0.25e when sampling in counting mode. (Please
re-input the sample number)
5. UPDATA Inputted sample number is zero when sampling in counting mode. (Enter
the counting mode after the UPDATA displayed for 1 second, operate
according to the previous sampling result, do not input 0 as the sample
number again)
6. bAt-Lo Low voltage for the battery (please recharge)
10