69
IM 04L41B01-02E
Setup Items and
Default Values
Environment > Security, Media save
Setup Item Selectable Range or Selections Default Value Setting
Security > Key Off/Login/Keylock Off
Security > Communication Off/Login Off
Save > Auto save On/Off On
Environment > Batch
Setup Item Selectable Range or Selections Default Value Setting
On/Off On/Off Off
Lot-No. digit Off/4/6/8 6
Auto increment On/Off On
Environment > Service port
Setup Item Selectable Range or Selections Default Value Setting
FTP 0 to 65535 21
HTTP 0 to 65535 80
SNTP 0 to 65535 123
Modbus 0 to 65535 502
Environment > Math
Setup Item Selectable Range or Selections Default Value Setting
Value on Error +Over/–Over +Over
Value on Overflow > SUM, AVE Error/Skip/Limit Skip
Value on Overflow > MAX, MIN, P-P Over/Skip Over
Environment > Report
Setup Item Selectable Range or Selections Default Value Setting
Report select > 1 Max/Min/Ave/Sum/Inst Ave
Report select > 2 Off/Max/Min/Ave/Sum/Inst Max
Report select > 3 Off/Max/Min/Ave/Sum/Inst Min
Report select > 4 Off/Max/Min/Ave/Sum/Inst Sum
File type Separate/Combine Separate
Keylock
Setup Item Selectable Range or Selections Default Value Setting
Password 8 characters or less –
Key action > START Free/Lock Free
Key action > STOP Free/Lock Free
Key action > MENU Free/Lock Free
Key action > USER Free/Lock Free
Key action > DISP/ENTER Free/Lock Free
Key action > FAVORITE Free/Lock Free
Media > External media Free/Lock Free
Action of Function > AlarmACK Free/Lock Free
Action of Function > Message/Batch Free/Lock Free
Action of Function > Math Free/Lock Free
Action of Function > Data save Free/Lock Free
Action of Function > E-mail/FTP Free/Lock Free
Action of Function > Time set Free/Lock Free
Action of Function > Display Function Free/Lock Free
Setup Items and Default Values