EasyManua.ls Logo

YOKOGAWA FLEXA SENCOM FLXA202 - App. 3.3 CSA FLXA21: Intrinsic Safety, Nonincendive

YOKOGAWA FLEXA SENCOM FLXA202
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<Appendix 3 Control Drawings>
App.3-6
IM 12A01A02-71E 3rd Edition : Mar. 23, 2018-00
App. 3.3 CSA
FLXA21: Intrinsic safety, Nonincendive
Model: FLXA21 / FLXA202 Date: May 29, 2017
Rev. Doc. No.: ICS032-A72 P.1
Yokogawa Electric Corporation
Control drawing (FOUNDATION Fieldbus / PROFIBUS PA type)
Installation for Zone 0, 1 / Division 1
Applicable models: FLXA21-D-x-x-CD-xx-xx-F-..., FLXA21-D-x-x-CD-xx-xx-P-…
Non-hazardous Area Hazardous
rea
Class I, Zone 0, 1, Group IIC, or
Class I, Division 1, Groups A, B, C, D
Temperature Class: T4
Measuring Module 1 (Note 6):
Type of Measuring Module
pH, SC, DO ISC SENCOM
Uo 11.76 V 11.76 V 5.36 V
Io 116.5 mA 60.6 mA 106.16 mA
Po 0.3424 W 0.178 W 0.1423 W
Co 100 nF 100 nF 31 μF
Lo 1.7 mH 8 mH 0.45 mH
Specific condition of use
- Electrostatic charges on the non-metallic or coated parts of the two wire analyzer shall be
avoided.
Supply +
+
A
ssociated Apparatus
(Note 3)
Supply
Housing Assembly
Measuring Module 1
FLXA21 Analyzer
Sensor 1
(Note 6)
T
T
Terminators
Other
Field Devices
Supply +, Supply
(Note 3):
FISCO field device
or
Ui: 24 V
Ii: 250 mA
Pi: 1.2 W
Ci: 2.72nF
Li: 0 mH

Table of Contents

Related product manuals