CALIBRATION
6-7. Temperature calibration
In order to make the most accurate measure-
ments, it is important to have a precise temper-
ature measurement. This affects the display of
temperature, and the output signal when used.
More important, however, is the temperature
compensation, and calibration accuracy.
The temperature of the sensor system should
be measured independently with a high preci-
sion thermometer. The display should then be
adjusted to agree with the reading (zero offset
calibrationonly).Forbestaccuracythisshould
be done as near to the normal operating tem-
perature as possible.
Press the keyandchoose“
Execute: Cali-
bration.
” Press “
Temperature calibration
” to
execute it.
6-8. Operation of hold function during cali-
bration
EXAxtISC450GhasaHOLD function that will
suspend the operation of the control/alarm re-
lays and mA-outputs.
During calibration, the user may choose to
enable HOLD so that the output signals are
frozentoa“
last
”or“
fixed
”value.Someusers
willchoosetoleavetheoutputs“
live
”torecord
the calibration event. This has implications
for pharmaceutical manufacture, for example,
where an independent record of calibrations is
mandatory.
Press to select Execute: “
HOLD
” and then
choose “
Manual Hold ON
” or “
Manual Hold
OFF
”.Thisallowsyoutosetupmanualhold.
Tocancelmanualhold,pressthelitHOLDsec-
tion on the Main display.
Press “
HOLD
” button on mainscreen, to re-
movetheHOLD.
TherouteforHOLDsetupis
Commissioning >> Output setup>> Config-
ure Hold